Your position: Home  ≡   Products  ≡   7-Series Sensors  ≡   7H2S-200 Sensor

7H2S-200 Sensor

Update time:Jun-06-15

The sensor is designed for the measurement of H2S concentration in gas phase. It can be used as the pin to pin replacement of the standard 7 series electrochemical H2S sensor.
Performance Characteristics
Nominal Range 0~200 ppm
Maximum Overload 1000 ppm
Sensitivity(20℃) 0.37 ± 0.07μA/ppm
Response Time (T90) ≤ 30 s
Zero Signal(20℃) < ± 0.4μA
Baseline Shift (- 40℃ ~ 50℃) < 0.5 ppm
Resolution 0.25 ppm
Linearity linear up to 200ppm
Bias Voltage zero
Environmental
Temperature Range -40℃ ~ 50℃
Pressure Range 1 atm ± 10%
Humidity Range 15% ~ 95% RH non-condensing
Life time
Long time output drift < 2% signal/month
Recommended Storage Temp 10℃ ~ 30℃
Expected Operating Life 2 years in clean air
Storage Life 6 months in original packaging
Warranty 12 months
Intrinsic safety Data
Maximum Current at 1000ppm H2S < 0.5mA
Maximum O/C Voltage 1.3V
Maximum S/C Current < 1.0A
Mechanical
Housing Material ABS
Weight (Nominal) 8g
Orientation None