Your position: Home  ≡   Products  ≡   4-Series Sensors  ≡   4ETO-100 Sensor

4ETO-100 Sensor

Update time:Jun-06-15

The sensor is designed for the measurement of Ethylene Oxide (ETO) concentration in gas phase. It can be used as the pin to pin replacement of the standard 4 series electrochemical Ethylene Oxide sensor.
Performance Characteristics
Nominal Range 0~100 ppm
Maximum Overload 100 ppm
Sensitivity(20℃) 0.25 ± 0.125μA/ppm
Response Time (T90) ≤ 120 s
Zero Signal(20℃) -0.2~2.5μA
Baseline Shift (- 20℃ ~ 50℃) < 10 ppm
Resolution 1 ppm
Linearity linear up to 100ppm
Bias Voltage +300 mV
Environmental
Temperature Range -20℃ ~ 50℃
Pressure Range 1 atm ± 10%
Humidity Range 15% ~ 90% RH non-condensing
Life time
Long time output drift < 2% signal/month
Recommended Storage Temp 10℃ ~ 30℃
Expected Operating Life 3 years in clean air
Storage Life 6 months in original packaging
Warranty 12 months
Intrinsic safety Data
Maximum Current at 100ppm ETO < 0.2mA
Maximum O/C Voltage 1.3V
Maximum S/C Current < 1.0A
Mechanical
Housing Material ABS
Weight (Nominal) 5g
Orientation None